China Suppliers Factory Pulmonary Function Detector LUD-V2 for Accurate COPD Asthma Screening and Respiratory Health Testing
Product Video
Product Parameters
- 1. Measurement Method: Differential pressure flow test method
- 2. Sensor: High precision double differential pressure flow sensor
- 3. Flow Measurement Range: 0 L/s ~ ±14 L/s
- 4. Flow Measurement Error: Maximum allowable flow error: ±10 L/min (±0.17 L/s) or ±5% of the reading
- 5. Volume Measurement Range: 0–8 L | Volume measurement error: ±3% or ±0.050 L, whichever is the maximum
- 6. Airflow Resistance: Should not exceed 0.35 KPa/(L/s) within the test range
- 7. Frequency Response: No more than 0.25 L/s or 12%
- 8. Linearity: Difference of average error of two adjacent test flows does not exceed 5%; test flow change does not exceed 0.17 L/S or 5% of the reading
- 9. Pulmonary Function Measurement: Forced Vital Capacity (FVC), Slow Vital Capacity (SVC), and Voluntary Ventilation (MVV)
- 10. Guided Measurement: Lung function measurements guided by animation
-
11. Measurement Parameters:
FVC, FEV0.5, FEV1, FEV3, FEV1/FVC, PEF, MEF75, MEF50, MEF25, MMF, OI, FIVC, PIF, FIV1; VC, TV, ERV, IRV, IC, VC/HT; MVV, RR, MVV/BSA, TV, BSA - 12. Curves Generated: Volume-flow velocity curve and time-volume curve based on pulmonary function measurement results
- 13. Filter & Mouth Design: Unique design for tighter seal, preventing air leakage during measurement
- 14. Assessment Functions: COPD assessment and ventilation defect assessment
- 15. System Software Functions: Instrument zero & calibration, doctor management system, history review system, medical record review, user file management
- 16. Display: ≥14.1 inch high-definition color display
- 17. Host Structure: Built-in industrial control host for more stable performance
- 18. Input Device: Logitech wireless keyboard and mouse for convenient operation
- 19. Design: Portable horizontal structure, exquisite and beautiful
- 20. Service Life: 8 years
- 21. Operating System: WIN7 32-bit / 64-bit, WIN10 64-bit compatible
- 22. Power Input: AC 220V, Current 0.75A, Power ≤60VA
- 23. Dimensions & Weight: Depth 226mm × Width 400mm × Height 353mm | Weight: ≤8 kg
After-Sales Service
-
🛡️One Year Warranty — Full warranty coverage provided for one year from the date of purchase.
-
🔧Permanent Maintenance Service — Outside the warranty period, only materials and maintenance fees will be charged; no labor fees will be applied.
-
⚡Rapid Response Guarantee — The manufacturer provides 2-hour immediate response and 48-hour fault resolution service.
Frequently Asked Questions
Q
What measurement method does this pulmonary function testing device use?
This device uses the differential pressure flow test method, equipped with a high-precision double differential pressure flow sensor. The flow measurement range covers 0 L/s to ±14 L/s, ensuring accurate and reliable results across a wide range of patients.
Q
Which pulmonary function tests can this device perform?
The device supports Forced Vital Capacity (FVC), Slow Vital Capacity (SVC), and Maximum Voluntary Ventilation (MVV). It also measures a comprehensive set of parameters including FEV0.5, FEV1, FEV3, PEF, MEF75, MEF50, MEF25, MMF, and more. COPD assessment and ventilation defect assessment are also supported.
Q
What operating systems are compatible with this device?
The device is compatible with Windows 7 (32-bit and 64-bit) and Windows 10 (64-bit). It features a built-in industrial control host for stable performance and comes with a Logitech wireless keyboard and mouse for convenient daily operation.
Q
What is the warranty policy and after-sales service for this product?
The product comes with a one-year warranty. After the warranty period, permanent maintenance service is available — only materials and maintenance fees are charged, with no labor fees. The manufacturer guarantees a 2-hour immediate response and commits to resolving faults within 48 hours.
Q
How does the device ensure measurement accuracy and prevent air leakage?
The device features a unique filter and mouthpiece design that provides a tighter seal to prevent air leakage during measurement. The maximum allowable flow measurement error is ±10 L/min (±0.17 L/s) or ±5% of the reading, and volume measurement error is ±3% or ±0.050 L, ensuring clinical-grade accuracy.
Q
What are the physical dimensions and power requirements of this device?
The device measures 226mm (D) × 400mm (W) × 353mm (H) and weighs ≤8 kg, making it a compact and portable horizontal-structure unit. Power input requires AC 220V, 0.75A, with total power consumption of ≤60VA. The display is ≥14.1 inches with high-definition color output.